Ion Selective Electrodes for Potentiometric Determination of Cyclizine in Its Pharmaceutical Dosage Form

×

Error message

User warning: The following theme is missing from the file system: journalijdr. For information about how to fix this, see the documentation page. in _drupal_trigger_error_with_delayed_logging() (line 1138 of /home2/journalijdr/public_html/includes/bootstrap.inc).

International Journal of Development Research

Volume: 
08
Article ID: 
11595
5 pages
Research Article

Ion Selective Electrodes for Potentiometric Determination of Cyclizine in Its Pharmaceutical Dosage Form

Seema Rani and Sarma,B.K.

Abstract: 

The two ion-pair complex of Cyclizine (Cy) with sodium tetraphenyl borate (TPB) and Phosphotungstic acid (PT) was prepared and used for the construction of Cy-selective electrode. The electrode based on the Cy-PT ion-pair complex has a lower detection limit of 1.5 x 10-7 M in a linear concentration range of 3.5 x 10-7 – 1.0 x 10-2 M with a slope of calibration curve of 50.5 mV/decay. The selectivity coefficient was calculated with separate solution method dictates the high selectivity of the electrode over other tested ions.

Download PDF: